Select your language

Photometry

  • Photometry

    Ruby transmittance
    Von FDominec, CC BY-SA 4.0    
    UV-VIS-NIR-Transmissionsspectrum
    of a one cm thick Rubin-Cristall

    Photometry

    Context: in aqua- and hydroponics there is no getting around measurements of nutrients as well as "pollutants". Photometry is the cheapest and most precise solution to this problem. It is actually only useful for professional use, as the acquisition costs are between 2,500 and 16,000 euros. A suitable device such as the Hach DR 6000 photometer costs about 13,000 euros (as of 2022-10). The cost of a measurement is between 2 and 15 euros, depending on the substance. However, this device is one of the most sophisticated measuring devices, it is already considered a "Porsche" among measuring devices - especially in terms of the speed of analysis.
     

     
    Photometry refers to all light-based measurement procedures that are carried out with a photometer (a light source with clearly defined values).
    photometer (a light source with clearly defined values).
    Photometry is based on the principle that each dye absorbs specifically at a certain wavelength depending on the concentration and the layer thickness.This relationship is described in Lambert-Beer's law.

     

     

     
    A photometer always has the same structure:
     
    • Light source
    • Monochromator
    • Sample in cuvette
    • Detector

    The sample
    The light now passes through our sample with a certain initial intensity I0. This is in a cuvette. A cuvette is a sample container that is transparent to allow the measurement. Substances always appear in the complementary colour to the absorbed colour. There are the following requirements for the sample:
     
    • The solution containing the sample must be homogeneous: it must be clear and not milky.
    • The sample should absorb light at the measured wavelength.
    • The concentration should be low, because at high concentrations Lambert-Beer's law no longer applies.
    • Now the light passes through the sample, loses intensity and therefore only has intensity I.

     

     

    Extinction as a central value

    What is extinction?

    The absorbance is the decadal logarithm of the ratio of the initial intensity and the intensity after sampling. The concentration can be determined by the loss of intensity.

    Extinction should not be confused with absorption. Extinction includes all light-attenuating events. The following light-attenuating events can occur in our sample:

    • The absorption of the wavelength by the molecules of the sample in the solution,
    • the refraction of light in an inhomogeneous, milky solution on the particles of the sample,
    • the reflection on the liquid surface or the cuvette.

    Preparation of the sample

    In order to specifically measure absorption, the following things must be taken into account:

    • The sample must be well dissolved: There should be no more particles floating around that would make the sample milky or inhomogeneous.
    • A calibration measurement must be carried out with the cuvette and the solvent. This means that the solvent (usually water) and the cuvette must be placed in the photometer and calibrated. (Press the Calibrate button).

    The photometer now measures the extinction again. However, since no dye is included, only the reflection on the water and on the cuvette is measured, which is subtracted from the extinction in the subsequent measurements. If these steps are followed, you can successfully measure the absorbance together with the extinction.

    Photometry and Lambert-Beer's law

    What does extinction tell us? Lambert-Beer's law applies here. This represents the extinction in connection with our substance, its concentration and the layer thickness of the optical medium. The layer thickness is, so to speak, the width of the cuvette, which is usually standardized to 1 cm. If the layer thickness remains constant, as does the molar, decadal extinction coefficient (absorption of the corresponding wavelength), there is a linear function here. This line increases as the concentration of the sample in the solution increases.

    This allows the concentration of the substance being sought to be determined. Note: Lambert-Beer's law only applies at low concentrations of the "dye". Thus there is a natural upper limit for the extinction. As an example, let's assume our solution is completely black, so no more light comes through. If you add more dye, the concentration increases, but no light comes through the solution for measurements anyway.

    In reality, there is no material that exactly fulfills Lambert's law. In particular, the radiance of any surface has a directional dependence and this changes as the direction from which the surface is illuminated changes. Even standards that are used to calibrate measuring devices can only be well described by Lambert's law in certain reflection directions and wavelength ranges. At wavelengths outside the visible spectral range and at reflection or illumination directions of more than a few 10° to the vertical, deviations of several 100% from Lambert's law can occur, even with normal ones. [1]

     

    Creating a spectrum with a photometer

    As a rule, to determine the concentration of an analyte in the sample, it is sufficient to measure the extinction or absorption of the solution. However, if you want to characterize your analyte more precisely, you may have to record a spectrum.

    As a rule, the extinction is measured individually for each wavelength at a specified concentration and layer thickness. However, since one does not want to carry out a calibration for every wavelength, modern spectrometers are used today that take on this task independently. Spectral analysis is an important method for identifying and/or determining the concentration of unknown substances.

     

    UV/VIS spectroscopy

    When we talk about spectra, we are no longer talking about photometry, but rather about spectroscopy or, more precisely, UV/VIS spectroscopy. The name comes from the fact that these spectra are recorded from the UV (Ultra Violet) to the visible (Visible) range of light.

    UV/VIS spectroscopy is based on measuring the absorbance of visible and ultraviolet light by the sample. The spectral, ie wavelength-dependent, information can be obtained either by selecting and scanning the wavelength of the incident light in front of the sample (see dual-beam spectrometer) or by separating the wavelengths of the transmitted light after the sample (diode array spectrometer). The ratio of the spectral intensity of the transmitted and incident light provides the transmission spectrum. The logarithmic reciprocal of the transmission gives the extinction spectrum.

    Basically, extinction provides information about absorption, scattering, diffraction and reflection on and in the sample. Phenomena of radiation absorption are often evaluated in UV/VIS spectroscopy, since the photon energy of visible and ultraviolet light corresponds to the transition energy of the states of outer electrons of many atoms and molecules. By absorbing photons in the visible and ultraviolet spectral range, valence electrons (e.g. those in the p and d orbitals) can be excited, that is, transformed into a state of higher energy. The transmission or extinction spectrum therefore allows the identification and quantitative determination of analytes. 

    Here is an overview of spectrophotometers and HowTo's.


    [1]  Andreas Höpe, Kai-Olaf Hauer: Three-dimensional appearance characterization of diffuse standard reflection materials. In: Metrologia. Band 47, Nr. 3, April 2010, S. 295–304, doi:10.1088/0026-1394/47/3/021.

    Kontext:  
    ID: 

     
     
  • Spectrophotometry: Methods and wavelengths

    Spectrophotometry is a technique used to measure the intensity of light absorption or transmission by a substance as a function of wavelength. It is widely employed in various scientific fields, including chemistry, biochemistry, molecular biology, environmental science, and analytical chemistry. Spectrophotometry allows researchers to quantitatively analyze the concentration of substances, identify compounds, study reaction kinetics, and assess the purity of samples.

    There are two main types of spectrophotometry methods:

    UV-Visible Spectrophotometry:

  • UV-Visible spectrophotometry measures the absorption of ultraviolet (UV) and visible (Vis) light by a substance.
    • In this method, a spectrophotometer emits a broad spectrum of UV and visible light through a sample, and the intensity of transmitted or absorbed light is measured by a detector.
    • The Beer-Lambert law is commonly used to relate the absorbance of light by a sample to its concentration and the path length of the light through the sample.
    • UV-Vis spectrophotometry is widely used in biochemistry for analyzing nucleic acids, proteins, and other biomolecules, as well as in chemical analysis for determining the concentration of various compounds.
      Infrared (IR) Spectrophotometry:
    • Infrared spectrophotometry measures the absorption of infrared radiation by a substance.
    • This method is based on the principle that molecules absorb infrared radiation at specific wavelengths corresponding to the vibrational and rotational modes of the chemical bonds within the molecule.
    • IR spectrophotometry is used for structural analysis, identification of functional groups, and qualitative and quantitative analysis of organic and inorganic compounds.

    Spectrophotometers are equipped with monochromators or filters to select specific wavelengths of light for analysis. The choice of wavelength depends on the nature of the sample and the information sought. For example:

      UV-Visible Wavelengths:
    • UV region: Typically ranges from 200 to 400 nm. Common applications include DNA and protein analysis, enzymatic assays, and measurement of organic compounds.
    • Visible region: Ranges from 400 to 800 nm. Used for colorimetric assays, analysis of pigments, and determination of metal ions.
    • Infrared Wavelengths:
    • Near-infrared (NIR) region: Ranges from 700 to 2500 nm. Used for analysis of organic functional groups, polymers, and pharmaceuticals.
    • Mid-infrared (MIR) region: Ranges from 2500 to 25000 nm. Suitable for identifying organic compounds, measuring bond vibrations, and characterizing materials.

    Overall, spectrophotometry is a versatile analytical technique with applications across various disciplines, offering valuable insights into the properties and composition of substances through the measurement of light absorption at specific wavelengths.


    Photometric methods    
    parameter method λ (nm)
    Alkalinity Bromocresol green 610
    Alkalinity, sea water Bromocresol green 610
    Aluminum Aluminum 530
    Brmine DPD 525
    Calcium Bromocresol green 466
    Calcium, sea water Zinkon method 610
    Chlorine, free, HK DPD 525
    Chlorine dioxide Chlorophenol red 575
    Chloride Mercury(II) thiocyanate 455
    Chromium (VI), HK Diphenylcarbohydrazide 535
    Chromium (VI), NK Diphenylcarbohydrazide 535
    Chrome, total and VI, 16 mm cuvette Diphenylcarbohydrazide 525
    CSB, HK EPA USEPA 410.4 610
    COD, NK EPA* USEPA 410.4 420
    Cyanide Pyridine Pyrazalone 610
    Cyanuric acid Turbidimetric 525
    Iron, NK TPTZ 575
    Color measurement Platinum cobalt 460
    Fluoride, NK SPADNS 575
    Total ammonium, NK Nessler 425
    Total hardness, NK EPA 130.1 466
    Hardness calcium Kalmagite 523
    Hardness magnesium EDTA 523
    Hydrazine p-Dimethylaminobenzaldehyde 466
    Iodine DPD 525
    Potassium, NK Turbidimetric, tetraphenyl borate 466
    Silica, NK Molybdenum blue 610
    Copper, NK EPA method 575
    Magnesium Kalmagite 466
    Manganese, NK PAN 560
    Molybdenum Mercaptoacetic acid 420
    Nickel, NK PAN 565
    Nitrate (NO3--N nitrate-nitrogen)**** Chromotropic acid 410
    Nitrate (NO3--N nitrate-nitrogen)**** Cadmium reduction 525
    Nitrite (NO2-) HK Iron sulfate 525
    Nitrite (NO2--N nitrite nitrogen), MK, 16 mm cuvette Diazotization 525
    Nitrite (NO2--N nitrite nitrogen), NK Diazotization 480
    Nitrite (NO2--N nitrite nitrogen), NK, 16 mm cuvette Diazotization 525
    Nitrite, sea water, UNK (nitrite nitrogen) Diazotization 480
    Ozone DPD 525
    pH Phenol red 525
    Phosphate, HK amino acid 525
    Phosphate, NK Ascorbic acid 610
    Phosphorus reactive, HK Vanadomolybdophosphoric acid 420
    Phosphorus, reactive, NK Ascorbic acid 610
    Phosphorus acid hydrolyzable Ascorbic acid 610
    Total phosphorus, NK Ascorbic acid 610
    Phosphorus, total HK Vanadomolybdophosphoric acid 420
    Oxygen, dissolved Winkler 466
    Oxygen scavenger (hydroquinone) Iron reduction 575
    Oxygen scavenger (carbohydrazide) Iron reduction 575
    Oxygen scavenger(DEHA) Iron reduction 575
    Oxygen scavenger (iso-ascorbic acid) Iron reduction 575
    Silver PAN 570
    Nitrogen, total, NK Chromotropic acid 420
    Sulfate Turbidimetric 466
    Surfactants (anionic) USEPA 425.1 610
    Surfactants (anionic) (SDBS), 16 mm cuvette Methylene blue 610
    Surfactants (non-ionic) (TRITON X-100), 16 mm cuvette TBPE 610
    Zinc Zinkon method 620
    * Dichromat EPA** Dichromat ISO 15705:2002
    *** Dichromate mercury-free for chloride-free samples
    **** Corresponds to a measurement range of 0 to 100 mg/L as NO3-† Whichever is greater
    HK = High concentration range
    MK = Medium concentration range
    NK = Low concentration range
    UHK = Ultra High Concentration Range
    UNK = Ultra-low concentration range

    Context: 

    ID: 147